Carefully read the centrifuge instruction manual and the rotor instruction manual before using the rotor.

T12A31 Fixed angle rotor 

  • Max. speed 12,000 rpm
    Max. R.C.F. 14,100 × g *
    Rotor nominal capacity 0.2 mL × 48 = 9.6 mL
    k factor 253
    Weight 1.6 kg

    *This is the value of the corner tube cavity. The value of the center tube cavity is 11,600xg.

 View of the size

(1)Summary of tubes *1 *2

Tubes & Bottles & Microplates Adapter & Bucket Cap Tube
position
Maximum
Speed
(rpm)
Maximum
R.C.F.
(xg)
Nominal
Vol.
(mL)
Part No.
Name
Size
(cm)
Material Net Vol.
(mL)
Part No.
Name
Size
(cm)
Material Number
of tubes
Part No.
Name
0.2 Commercial item
0.2mlPCR Tube
φ0.6×2.4 --- 0.2 *3 --- --- --- --- --- outside *4 12,000 14,100 *3
--- --- --- --- --- inside1 *4 12,000 11,600 *3
Commercial item
0.2ml PCR Tube (8-tube strip)
--- --- 0.2 *3 --- --- --- --- --- outside *4 12,000 14,100 *3
--- --- --- --- --- inside1 *4 12,000 11,600 *3
Commercial item
0.2ml PCR Tube (12-tube strip)
--- --- 0.2 *3 --- --- --- --- --- outside *4 12,000 14,100 *3
--- --- --- --- --- inside1 *4 12,000 11,600 *3

*1:No tubes come with the rotor. Please purchase your required tubes separately.

*2:Please refer to the brochure of each commercially available tube and/or bottle for its actual volume.

*3:The indicated maximum R.C.F. value is based on the value of Rmax indicated in brackets. The actual maximum R.C.F. value of the tube may be different.

*4:“Outside” means the corner of the tube cavities. “Inside 1” means the center of the tube cavities.

※When using commercially available tubes and/or bottles, perform centrifugations under the allowable speed or allowable RCF specified by their manufacturer. Otherwise, the tubes and/or bottles may be damaged during centrifugations.