Carefully read the centrifuge instruction manual and the rotor instruction manual before using the rotor.

R6S Swinging bucket rotor 

  • Max. speed 5,700 rpm
    Max. R.C.F. 5,010 × g
    Rotor nominal capacity ※Please inquire.
    k factor 648
    Weight 10.7 kg
 View of the size

(1)Summary of tubes *1 *2 *3

Tubes & Bottles & Microplates Adapter & Bucket Cap Maximum
Speed
(rpm)
Maximum
R.C.F.
(xg)
Nominal
Vol.
(mL)
Part No.
Name
Size
(cm)
Material Net Vol.
(mL)
Part No.
Name
Size
(cm)
Material Number
of tubes
Part No.
Name
2 Commercial item
2.0ml Micro Tube
φ1.08×4.2 PP 2 S408787A
(2pcs/set)
2×48S Adaptor
--- PP --- 5,700 5,010
1.5 Commercial item
1.5ml Micro Tube
φ1.08×3.9 PP 1.5 S408787A
(2pcs/set)
2×48S Adaptor
--- PP --- 5,700 5,010
0.5 Commercial item
0.5ml Micro Tube
φ0.78×3.0 PP --- S408787A
(2pcs/set)
2×48S Adaptor
--- PP --- 5,700 5,010
0.2 Commercial item
0.2mlPCR Tube
φ0.6×2.4 --- 0.2 S408786A
(2pcs/set)
0.2×96MP Adaptor
--- PP 1 --- 5,700 5,010
Commercial item
0.2ml PCR Tube (8-tube strip)
--- --- 0.2 S408786A
(2pcs/set)
0.2×96MP Adaptor
--- PP 1 --- 5,700 5,010
--- Commercial item
96-hole Microplate for PCR (V-bottom)
12.8×8.6×1.4 --- --- S408786A
(2pcs/set)
0.2×96MP Adaptor
--- PP 1 --- 5,700 5,010
Commercial item
96-hole Microplate *4
12.8×8.6×1.4 --- --- --- --- --- --- --- 5,700 5,010

*1:In case the weight of contents in the bucket exceeds 400g, decrease the speed according to the instruction manual.

*2:For commercially available tubes, some of them may not withstand the centrifugation at the maximum rotational speed. Conducting a preliminary experiment is highly recommended before the centrifugation with the actual sample. The preliminary experiment means to centrifuge the tube filled with water at the maximum rotational speed and check if there is any abnormality on the tube.

*3:No tubes and/or microplates come with the rotor. Please purchase your required tubes/microplates separately.

*4:Please use the adapter (standard accessory) to use commercially available 96-hole microplate.

※When using commercially available tubes and/or bottles, perform centrifugations under the allowable speed or allowable RCF specified by their manufacturer. Otherwise, the tubes and/or bottles may be damaged during centrifugations.

(2)Standard accessories

Part No. Name Q’ty
--- Adapter 2

(3)Optional accessories

Please use following parts to maintain the rotor.

Part No. Name Q’ty
84810601 Lubricate for Screw 1