Carefully read the centrifuge instruction manual and the rotor instruction manual before using the rotor.
T15AP31(T15AP21) Fixed angle rotor
-
Max. speed 15,000 rpm Max. R.C.F. 18,800 × g Rotor nominal capacity 2 mL × 18 = 36 mL k factor 678 Weight 0.7 kg

(1)Summary of tubes *1 *2
Tubes & Bottles & Microplates | Adapter & Bucket | Cap | Maximum Speed (rpm) |
Maximum R.C.F. (xg) |
|||||||
---|---|---|---|---|---|---|---|---|---|---|---|
Nominal Vol. (mL) |
Part No. Name |
Size (cm) |
Material | Net Vol. (mL) |
Part No. Name |
Size (cm) |
Material | Number of tubes |
Part No. Name |
||
2 |
Commercial item
2.0ml Micro Tube |
φ1.08×4.2 | PP | 1.7 | --- | --- | --- | --- | --- | 15,000 | 18,800 *3 |
1.5 |
Commercial item
1.5ml Micro Tube |
φ1.08×3.9 | PP | 1.4 | --- | --- | --- | --- | --- | 15,000 | 18,800 *3 |
1 |
489241A (100pcs/set) 1.0ml Round-bottom Glass Tube |
φ0.9×3.5 | Glass | 0.7 |
S413789A (2pcs/set) 1GA2 Adaptor |
--- | PP | 1 | --- | 15,000 | 18,800 *3 |
0.7 |
Commercial item
0.7ml Micro Tube |
φ1.08×2.5 | PP | 0.6 | --- | --- | --- | --- | --- | 15,000 | 18,800 *3 |
0.5 |
489240A (100pcs/set) 0.5ml Glass Tube |
φ0.9×3.5 | Glass | 0.5 |
S413789A (2pcs/set) 1GA2 Adaptor |
--- | PP | 1 | --- | 15,000 | 18,800 *3 |
489239A (100pcs/set) 0.5ml Round-bottom Glass Tube |
φ0.6×4.7 | Glass | 0.5 |
489238A (4pcs/set) 0.5 Glass Adaptor |
0.62×4.25 | ABS | 1 | --- | 15,000 | 18,800 *3 | |
Commercial item
0.5ml Micro Tube |
φ0.78×3.0 | PP | 0.5 |
487293A (4pcs/set) 0.75 Adaptor |
0.81×3.0 | ABS | 1 | --- | 15,000 | 18,800 *3 | |
0.4 |
Commercial item
0.4ml Micro Tube |
φ0.57×4.6 | PP | 0.4 |
487292A (4pcs/set) 0.4 Adaptor |
0.6×4.7 | ABS | 1 | --- | 15,000 | 18,800 *3 |
0.25 |
Commercial item
0.25ml Micro Tube |
φ0.57×3.0 | PP | 0.25 |
487292A (4pcs/set) 0.4 Adaptor |
0.6×4.7 | ABS | 1 | --- | 15,000 | 18,800 *3 |
*1:No tubes come with the rotor. Please purchase your required tubes separately.
*2:Please refer to the brochure of each commercially available tube and/or bottle for its actual volume.
*3:The indicated maximum R.C.F. value is based on the value of Rmax indicated in brackets. The actual maximum R.C.F. value of the tube may be different.
※When using commercially available tubes and/or bottles, perform centrifugations under the allowable speed or allowable RCF specified by their manufacturer. Otherwise, the tubes and/or bottles may be damaged during centrifugations.